Product Configurator

Manual Changeover Oxygen Delivery System for Ambulance

Continuous Oxygen Supply Systems for Ambulances with Manual Changeover Control between 2 sources

The Manual Changeover Oxygen Delivery System for Ambulances supports emergency medical services by delivering a steady oxygen supply to patients during transport to healthcare facilities.

The system manages two oxygen sources and allows medical staff to switch manually between the left and right sources for continuous supply.

A touchscreen alarm panel tracks gas pressure in every cylinder and alerts medical personnel when pressure drops. Oxygen outlets allow simple, safe connections and deliver a consistent, reliable oxygen supply to patients. A pressure gauge monitors line pressure to support optimal system performance.

Request a quote

Highlights

  • Manual switchover between left and right cylinders via a control knob ensures a consistent and dependable oxygen supply.
  • Medical Gas outlets type: Sil-Ohio, ENV 737-6, Puritan Bennett type, DIN, British standard, DISS*.
  • Control Alarm panel monitor highlights:
    • Presenting the actual time left for treatment (in minutes) based on preset flow rate parameters.
    • Visual & audio alarm signals in case of low pressure.
    • 10.1” full HD LCD touchscreen – graphically and digitally displaying the cylinder gas pressure, alerting in case of a pressure drop.
    • Periodical maintenance alert.
    • Gas expiration date alert – an alert for the medical gas cylinder expiration date.
    • Log History File- Event logging, recording all alarms and panel activities.
    • Customizable text & language.
    • A color trend arc showing the gas pressure status at each cylinder.
    • Automatic adjustment of screen brightness (day & night mode).
    • Advanced Communication: TCP/ IP Modbus.
  • SMS and email alert when oxygen pressure reduces to a pre-set level – optional.
  • Tailor-made installation.

 

Applicable Standards:

  • ISO 10524, EN-1789, NFPA 1917

Silbermann’s Quality Management System:

  • ISO 9001-2015
  • ISO 13485-2016

* The quantity and types of oxygen gas outlets and adapters are available based
on customer requirements.

Specifications

  • Gas sensors/Pressure 4-20 mA transmitters.
  • Oxygen Pressure Regulators**:
    • Input pressure: up to 200 bar
    • Output pressure: 3.5-4 bar (Regulated pressure)
    • Flow rate capacity of up to 100 liters per minute.
  • Flow selector 0-25 lpm – optional
  • Power Requirements:
    • Voltage: 12V DC supplied from the Ambulance
      Electrical Supply Power.
    • 5A Slow Blow fuse 5x20mm.

 

** Oxygen regulator with CGA 540 connector – optional

 

Contact Us

    We'd love to hear from you!

    #5 Harakevet st. POB 4605, Petah-Tikva ISRAEL
    +972-3-9309186 info@silbermnn.com