Product Configurator

Multi-Strecher Oxygen Delivery System

Central Oxygen Supply with Continuous Delivery and Automatic Changeover on the Go

The Multi Stretcher Oxygen Delivery System plays a critical role in emergency medical services by ensuring a continuous oxygen supply to multiple patients during transport to healthcare facilities.

Oxygen Delivery System with a two-source Manifold System

The system manages 2 sources and provides a continuous supply and a smooth automatic bank switchover by means of a pneumatic shuttle valve.
All medical gas information is displayed graphically and digitally on a 12” full HD LCD touch screen located on the wall inside the bus.
This touchscreen control panel monitors the gas pressure in each bank and shows/alerts the medical staff in case of a pressure drop.
Oxygen outlets are mounted on the interior wall of the bus, with two outlets positioned near each stretcher. enable easy and safe connection to the system, and ensure a constant and reliable oxygen flow for the patients, even during hours of high demand when all outlets are actively supplying oxygen.

Request a quote

Highlights

  • Continuous supply and smooth automatic bank changeover, without pressure fluctuation, by means of pneumatic shuttle valve.
  • Maintenance-friendly design – Each manifold, accommodating six 10L cylinders, ensuring easy access for maintenance and refilling.
  • LCD 12” touch screen display – Graphically & digitally display of the actual status of each source- “In use” or “Stand by”.
  • Visual & audio alarm signals in case of-change between banks, “Below min” and High/Low line supply pressure.
  • Medical Gas outlets type*: Sil-Ohio, ENV 737-6, Puritan Bennett type, DIN, BS, DISS, AGA.
  • Gas expiration date alert – an alert for the medical gas cylinder expiration date.
  • Leak detection from manifolds on Stand-By mode.
  • Log History – Event logging, recording all alarms and maintenance activities.
  • SMS alert when oxygen pressure reduces to a pre-set level – optional
  • Web remote view on the Internet – optional online view accessible via a web browser. This website allows authorized users to remotely access and view the data in real-time.
  • Additional data: GPS data showing the current location of each ambulance bus by its number. – optional
  • Consumption meter – optional

* Other types of oxygen gas outlets are available according to customers’ request.

System Components

  • 12” full HD LCD touchscreen.
  • up to 20 Medical Gas outlets**: Sil-Ohio, ENV 737-6, PB, DIN, BS, DISS, AGA
  • Gas sensors/Pressure 4-20 mA transmitters.
  • All materials are oxygen compatible and cleaned for oxygen use.
  • Fully Automatic Changeover unit including manual bypass.
  • Each cylinder station includes a HP valve, flexible stainless still pig tail with integrated NRV and a secure chain.
  • Oxygen High pressure regulator (primary regulators for manifold cylinders):
    • Input pressure: up to 200 bar
    • Output pressure: 11 bar (Regulated pressure)
    • Flow rate capacity of up to 100 liters per minute.
  • Oxygen Line pressure regulators (change over unit regulators):
    • Input pressure: 11 bar
    • Output pressure: 4 bar
  • When all gas outlets are actively delivering oxygen, the pressure remains stable between 3.5 – 4 bars.

Power Requirements

  • Voltage: 12V DC supplied from Ambulance Bus Electrical Supply Power.
  • 5A Slow Blow fuse 5x20mm.

Applicable Standards:

  • ISO 10524, EN-1789, NFPA 1917

Silbermann’s Quality Management System:

  • ISO 9001-2015
  • ISO 13485-2016

Contact Us

    We'd love to hear from you!

    #5 Harakevet st. POB 4605, Petah-Tikva ISRAEL
    +972-3-9309186 info@silbermnn.com